Model T940 User Manual
Publication No. 980938 Rev. K
Soft Front Panel Operation 5-24
Astronics Test Systems
Table 5-20: Record Mode Settings
Setting
Description
Typical Usage
Disabled
The contents of the record memory will not
change during the next burst if Step Record
Mode is set to either None or Record
Count.
Setting Record Mode to Disabled
insures that the record memory will not
be written to when Step Record Mode is
set to either None or Record Count.
This means that if errors were recorded
in a previous burst, they will remain in
memory throughout the current burst.
Non-
Error(0)
The contents of the record memory will be
set to 0 during the next burst if Step Record
Mode is set to either None or Record
Count.
Setting Record Mode to Non-Error(0)
when Step Record Mode is set to either
None or Record Count clears the record
memory during the next burst, insuring
that any previously recorded errors will
not persist.
The relevant VXI
plug&play
API function is:
•
tat964_setSequencerRecordMode
Config Data Sequencer A/B
The Configure Data Sequencer A/B panel is used to program the clock settings,
sequence control signals, timeout values, overcurrent, and record settings.
Access this panel from the menu bar:
Config > Data Sequencer x
. (Where “x”
is the sequencer you wish to configure.)
Figure 5-20: Configure Data Sequencer
Configure Clocks
Access this panel from the menu bar:
Config > Data Sequencer x> Clocks
.
(Where “x” is the sequencer you wish to configure.)