Model T940 User Manual
Publication No. 980938 Rev. K
Soft Front Panel Operation 5-166
Astronics Test Systems
Export
This command button saves the current calibration data to a comma separated
file with a format that can be loaded using the
File > Load DRA/DRB
Calibration
menu command.
The relevant VXI
plug&play
API functions are:
•
tat964_saveCalibrationFile
Stop
This command button stops a calibration or verification run.
Update
This command button writes the new calibration data to non-volatile memory.
The relevant VXI
plug&play
API function is:
•
tat964_updateCalibrationData
Monitor Temperature Panel
This panel shows the temperature of the following components within the DRM:
•
Digital Board Sequencer FPGAs
•
DR3e, DR9 and UR14 variable voltage pin electronics.
Trip Temperature
This control programs a trip point that will disconnect the power pins from the
variable voltage pin electronics and open the connect relays if the specified
temperature is exceeded. A Driver/Receiver temperature alert event is also
generated.
Figure 5-98: DB Monitor Temperature Panel