Publication No. 980938 Rev. K
Model T940 User Manual
Astronics Test Systems
Programmable Channel Calibration 6-9
2. Verify that the M ADC calibrate function is now in focus.
Select Start and End Channels and Measurement Delay
Equipment: Basic Setup
Procedure:
1. Select DRA or DRB (if installed) using the Driver/Receiver switch.
2. Use the Start and End Channel fields to select the I/O and Auxiliary
channels to be calibrated.
3. The minimum measurement delay for this calibration is 200 ms.
Increase this value to give the calibration points more time to settle.
DRM Calibration Warmup
Equipment: Basic Setup
Procedure:
1. Allow the T940 DRM to warm to its nominal application
temperature.
2. Hit the Continue button when the required temperature is reached.
If the temperature reaches 80°C, the process continues
automatically.