Publication No. 980938 Rev. K
Model T940 User Manual
Astronics Test Systems
Soft Front Panel Operation 5-97
Figure 5-54: Edit Vector Table Panel
Vector Bit Index
This allows the user to enter the index to program. There are 16 indexes that
can be set (0 to 15). The index is the binary value of the vector bits (VA0 through
VA3).
The relevant VXI
plug&play
API function is:
•
tat964_setVectorJumpTable
Vector Jump Step
This allows the user to enter the jump step number for the current vector jump
index.
The relevant VXI
plug&play
API function is:
•
tat964_setVectorJumpTable
Timing Set
When the timing mode is set to indexed, this control allows the user to specify
the timing set for the current vector jump index.
The relevant VXI
plug&play
API function is:
•
tat964_setVectorJumpTable
Set Channel Test
This command button displays the Edit Channel Test panel so the channel test
settings can be programmed for the selected sequencer.
Configuring the sequence channel test registers consists of the following:
1. Program the expect value
2. Program the mask value
The expect value is compared to the response high (Good 1) of the input
channel. A high in the mask, disables the comparison.
The result of all four channel test registers can be routed to the VXI TTL trigger