Page 32 · 6300 Series
ComboSource
User’s Manual
enough forward current through the laser diode to generate the selected monitor
diode current. Only low frequency modulation (10Hz or less) is possible in
Im
mode due to the feedback latencies of the photodiode itself.
Po
mode (also referred to as APC, or automatic power control, mode) is simply
Im
mode with a mathematical constant applied to the set point, providing a
convenient way of operating in milliwatts. Using the
PD Resp
factor (in
μ
A/mW),
a
Po
set point is internally converted to an equivalent
Im
set point by the driver,
which is then used to control the photodiode feedback. For example, if the
PD
Resp
factor was 10, then a set point of 1mW would be the same as a set point of
10
μ
A.
Vf
mode (also referred to as AVC, or automatic voltage control, mode) is used to
control the voltage driven through the device. Unlike ACC mode, AVC mode
allows the current to drive to whatever current is necessary to achieve the
voltage set point, so long as it does not exceed the current limit. As with
Im
mode, only low frequency modulation (10Hz or less) is possible in
Vf
mode.
Modulation
The instrument supports external analog modulation using the
Modulation
BNC
on the front panel of the instrument. Modulation rates vary by model, so see
your model’s specification for the maximum modulation rates. Only
Io HiBW
mode supports high speed modulation. All other modes of operation have a
modulation bandwidth of 10Hz or less.
Current Range
The
ComboSource
offers two current ranges: a high range, which is the
maximum rated current for the driver (e.g., 500mA on a
6305
) and low range,
which is one-half the maximum rated current for the driver (e.g., 250mA on a
6305
). The lower range offers improved accuracy (in
Io (ACC)
and
Io HiBW
modes) and improved stability in all modes.
Using the Laser Limits
Both the laser current and voltage limits are implemented in hardware, providing
for fast response to changes in laser diode operation. When a voltage limit is
detected, the output is immediately shutdown. Because of the sensitivity of the
voltage limit, operating near the limit (within one to two hundred millivolts) is not
recommended. In general, you should set the voltage limit to 0.1V to 0.2V higher
than any anticipated operating point. The voltage limit is tested against
the
voltage at the connector
, unless in remote voltage sense mode. Any
Laser
Cable R
value is ignored, as
Laser
Cable R
is a software only calculation, and
the voltage limit is implemented in hardware. See
Using the Cable R Setting
,
above, for more information on the
Laser
Cable R
setting.