Input Power Connector (IPC)
Once the module has been removed, remove the small, white voltage selection
tumbler from the module, and then re-insert so that the desired voltage is
shown. Re-insert the module into the IPC.
Powering Up the Unit
Once the correct voltage selection has been made, connect the AC power cord
to the unit. You must properly ground the unit by plugging the supplied power
cord into a three prong grounded outlet, or using a three-to-two prong adapter
and connecting the ground tab to earth ground.
Turn the power switch, located on the front panel, into the on (|) position. The
unit will display the model, serial number, and firmware version, go through a
quick power-up self-test, and return to the last known operating state.
Ventilation
The
ComboSource
has vent holes on both sides of the unit. You must not block
these vent holes, or overheating may occur, causing damage to the unit.
CAUTION
Do not exceed 250VAC on the line input.
It is critical to select the proper voltage selection prior to
applying power to the unit. If the actual voltage exceeds
the voltage selection by +/-10%, damage to the unit
may occur.