RS232 Connector
The RS232 connection is male DB-9 connector wired in a NULL modem
configuration.
Pin
Description
2
Receive
3
Transmit
5
Ground
1,4,6
Commoned together
7,8
Commoned together
9
No connection
Shell
Earth ground
RS232 Connector (DB-9 Male)
For more information on using the RS232 interface, see the
Computer
Interfacing Manual
which is included on the CD that accompanied this product.
Grounding Considerations
A key feature of the
ComboSource
is the optical isolation of both the
photodiode and modulation inputs. By isolating these inputs, earth grounding of
the photodiode anode or cathode, or earth grounding the modulation input,
cannot cause a ground loop through the instrument. Likewise, the laser anode
and cathode connections are also isolated from earth ground.
However, if you use the earth ground pin of the
Output
connector, it is possible
to create a ground loop if the instrument’s earth ground is connected to a fixture
or optical table that is also earth grounded. Make sure that from your laser diode
package there is only a single path to earth ground.
Remote Mode Operation
“Remote mode” operation is when the
ComboSource
is being controlled by a
computer over the USB or RS232 interfaces. When in remote mode, the
ComboSource
behaves differently, preventing you from affecting the operation
of the instrument, such as changing the set point. You cannot enter the menu
without taking the unit out of remote mode, and the knob is disabled, unless it
has been explicitly enabled by the
Set While Rmt
setting in the
Comm Menu
.
You can exit remote mode at any time by pressing the
Menu
button, which has
a secondary function to return the
ComboSource
to local operation.