6300 Series
ComboSource
User’s Manual · Page 11
Operation
The Front Panel
Operation of the
ComboSource
is very straightforward. The sections below will
help familiarize you with the front panel, the display, and the menu structure.
6300
ComboSource
Main Screen
The front panel is designed for ease of operation. There is an On/Off button for
the Laser and TEC outputs, one to change which set point is adjusted by the
knob, and one to enter the menu. The remaining three keys are the function
buttons, which can be assigned a variety of actions. More information on the
function keys are available in the section titled
Using the Function Keys
, below.
The outputs are controlled via the two
On/Off
buttons:
For the laser, when turning the output on, there is a programmable
delay (factory default of 3 seconds) where the unit will indicate that the
output is on (the blue output LED will be lit), but the output stage is, in
fact, not on. This is a CDRH safety feature required for some laser
applications, and this delay can be adjusted in the menu. The output
will be energized once the delay has expired.
For the TEC, when the output is turned on, it will immediately begin
driving current as needed to achieve the set point.
The
Menu
button is used to enter the
ComboSource
menu. When in remote
mode, it also acts as a
Local
button, returning the instrument to local mode.
The adjustment knob located on the right hand side of the unit is used to
change the set point or parameters in the menu. It also acts as a push/enter
Содержание ComboSource 6300 Series