6300 Series
ComboSource
User’s Manual · Page 57
disabled output
forced a shutdown of the laser output.
E-512
Power failure
A power failure was detected.
E-514
Laser mode change
disabled output
A change in the operating mode of the
ComboSource while the output was on shutdown
the output.
E-516
Incorrect configuration
for calibration to start
The ComboSource was not configured properly,
including the mode and output on state, to be
able to start the desired calibration process.
E-517
Calibration must have
the output on to start
The laser output must be on for the calibration
process to start.
E-521
TEC temperature limit
disabled output
The laser output was turned off because the TEC
temperature limit was exceeded corresponding
bit in the OUTOFF register was set.
E-534
Po mode selected with
PD Response set to
zero
Attempted to select Po mode and PD Response
was zero, or ComboSource was in Po mode and
PD Response was set to zero.
E-535
Calibration cancelled
The active calibration process was cancelled.
E-536
Intermittent contact
fault
The instrument detected an intermittent contact
and shut down the laser output. If this is
triggering falsely (such as in a noisy
environment), the intermittent contact detection
can be disabled in the main menu.
E-537
Thermal trip
Excessive power dissipated inside unit. Lower
voltage limit or add series resistance. See
“Thermal Considerations” for more details.
W-800
Remote voltage sense
is low
When in remote voltage sense mode, the
ComboSource detected a significant difference
between the remote voltage and the voltage at
the connector, which may indicate a problem with
the remote voltage sense connection.
W-801
Burst Mode, Hold
Output
When in Io (Burst) mode, to turn the output on,
the Output button must be held down for at least
one second. If it is held down for less than one
second, this warning message informs the user
than the Output button press did not turn the
output on.
E-998
Command not
supported
A command was recognized but not supported
by the ComboSource.
E-999
Non-specific error
A non-specific error was encountered.