Page 56 · 6300 Series
ComboSource
User’s Manual
E-410
Temperature was out
of tolerance, output
turned off
The temperature went out of tolerance and the
output was turned off.
E-415
Sensor short, output
turned off
A sensor short circuit was detected and the
output was turned off.
E-416
Cal failed
An error occurred during calibration that caused
the calibration to fail. The unit is no longer in
calibration mode.
E-419
TEC not stable
The TEC is considered stable if the temperature
has changed less than 0.02°C for more than 20
seconds.
E-433
Not a TEC
The TEC:CHAN command attempted to select a
non-TEC channel
E-434
Ite limit exceeds cable
rating
The cable plugged into the unit is not rated to
carry the current as set by the current limit. Use a
1262 cable, 1263 cable, or custom cable with ID
pin connected to analog ground, or reduce the
current limit below 5A.
E-435
Mode Change
A mode change occurred when the output was
on, forcing the output off.
E-436
AutoTune failed
The AutoTune process failed. See the “Using the
AutoTune Function” for more information.
E-437
AutoTune in T mode
only
The AutoTune process is only available in T
(temperature) mode. In R mode, the PID
parameters must be determined manually.
E-501
Interlock shutdown
output
The interlock input (pins 1 and 2 of the output
connector) were not shorted when the output was
on.
E-504
Laser current limit
disabled output.
The laser output was turned off because a current
limit was detected and the corresponding bit in
the OUTOFF register was set.
E-505
Laser voltage limit
disabled output
The laser voltage exceeded the voltage limit and
the output was turned off.
E-506
Laser photodiode
current limit disabled
output
The laser output was turned off because a
photodiode current limit was detected and the
corresponding bit in the OUTOFF register was
set.
E-507
Laser photodiode
power limit disabled
output
The laser output was turned off because a
photodiode power limit was detected and the
corresponding bit in the OUTOFF register was
set.
E-508
TEC off disabled
output
The laser output was turned off because the TEC
was off and the corresponding bit in the OUTOFF
register was set.
E-509
Laser short circuit
disabled output
The laser output was turned off because a short
condition was detected and the corresponding bit
in the OUTOFF register was set.
E-510
Laser out of tolerance
disabled output
The laser output was turned off because an out-
of-tolerance condition was detected and the
corresponding bit in the OUTOFF register was
set.
E-511
Laser control error
A hardware control error was detected which