![Array electronic 3720A Скачать руководство пользователя страница 64](http://html1.mh-extra.com/html/array-electronic/3720a/3720a_manual_2975639064.webp)
61
see 4.5.2 section for detailed operations)
2.000V CV OFF
Step2
Press
key to enter into the main menu.
Load Default
:
Yes *No
Step3
Use
key to select the menu item “CV Curr Limit”; use Entry
keys or the knob together with
,
keys to input the
current limit value to 20A. Press
key to confirm.
CV Curr Limit
:
20.000A
Step4
Press
key to exit the main menu.
0.000V 0.000A
2.000A CCH OFF
Step5
Press
key to turn on the load.
0.000V 0.000A
2.000A CCH ON
For the above operating procedures, the corresponding SCPI commands should be:
MODE
CV
;
Sets the mode
VOLT
2
;
Sets the current value to 2V
INP:LIM:CURR 20A
;
Sets the current limit value to 20A in CV mode
INP
ON
;
Turns on the load
4.14.5
Current Rise/Fall Rate in CC Mode
Example: Set the load to CCH mode, set the current value to 2A; set the current rise rate (Curr
Rise Rate) in CC mode to 0.002A/us; set the current fall rate (Curr Fall Rate) in CC mode to
0.005A/us; turn on the load; turn off the load. The detailed operations are shown below:
Procedures Operation
Descriptions
Display
1
Set the load to CCH mode, and set the current value to 2A.
0.000V 0.000A
2.000A CCH OFF
2
Press
key to enter into the main menu.
Load Default
:
Yes *No
3
Use
key to select the menu item “Curr Rise Rate”; use Entry
keys or the knob together with
,
keys to input the
current value to 0.002A/us. Press
key to confirm.
Curr Rise Rate
:
0.002A/us
4
Use
key to select the menu item “Curr Fall Rate”; use Entry
keys or the knob together with
,
keys to input the
current value to 0.005A/us. Press
key to confirm.
Curr Fall Rate
:
0.005A/us
5
Press
key to exit the main menu.
0.000V 0.000A
2.000A CCH OFF
6
Press
key to turn on the load.
0.000V 0.000A
2.000A CCH ON