CHAPTER 6
UPDATING THE SOFTWARE
1
Introduction
This section describes the procedure to upgrade the pump software to Version 3.01. The upgrade is
required to improve the reliability of the kinked-tube alarm for supine patients.
The upgrade should be carried out on any pump within the serial number range specified but only at the
following times:
•
At the next scheduled service.
•
After any repair carried out on the pump.
•
If the pump is exhibiting excessive kinked-tube alarms.
Serial Number range:
Before upgrading the pump to the new V3.01 software, check the current software version by initialising
the pump (refer to Chapter 6, Page 15, Section 8). If the software is already at V3.01 or higher, do not
attempt the upgrade.
To perform the check and the upgrade, a Personal Computer is required, equipped with a serial port and
running Microsoft
®
Windows
®
. This Laptop or PC must have the ArjoHuntleigh Serial Communications
Pack software installed in order to communicate with the pump. To install the Serial Communications
Pack, refer to Chapter 6, Page 9, Section 3.
Any pump up to and including serial number:
507032000
(a)
a. All pumps from serial number 507032001 onwards already have version 3.01 (or later)
software installed at the factory.
Содержание FLOWTRON UNIVERSAL
Страница 1: ...Service Manual FLOWTRON UNIVERSAL ...
Страница 2: ......
Страница 4: ...Flowtron Universal Service Manual Ammendment Record Issue 2 Page ii SER0009 Sep 2009 ...
Страница 8: ...Flowtron Universal Service Manual Issue 2 Page vi SER0009 Sep 2009 ...
Страница 22: ...Maintenance Flowtron Universal Service Manual Chap 3 Issue 2 Page 4 SER0009 Sep 2009 ...
Страница 38: ...Testing Flowtron Universal Service Manual Chap 4 Issue 2 Page 16 SER0009 Sep 2009 ...
Страница 88: ...Pump Repair Flowtron Universal Service Manual Chap 5 Issue 2 Page 50 SER0009 Sep 2009 ...
Страница 108: ...Updating the software Flowtron Universal Service Manual Chap 6 Issue 2 Page 20 SER0009 Sep 2009 ...
Страница 114: ...Parts List Flowtron Universal Service Manual Chap 8 Issue 2 Page 4 SER0009 Sep 2009 ...
Страница 116: ...SER0009_02 September 2009 ...