Testing
Flowtron Universal Service Manual
Chap 4
Issue 2
Page 2
SER0009
Sep 2009
2.3.2
Figure 14 shows a foot garment and a single-chamber calf or thigh/calf garment
connected to the pump.
Figure 14 - Foot and Single-Chamber Calf or Thigh/Calf Garments Connected
2.3.3
Figure 15 shows a foot garment connected to one hose set assembly, and no garment
connected to the other.
Figure 15 - Only One Garment Connected
2.3.4
If the LCD display shows either an incorrect garment graphic or a “no garment” graphic
when one is connected, then carry out the procedure for recalibrating the pump for a
new hose set assembly (Refer to Page 12, Section 9).
Table 3 - Garment Options and Default Therapy Parameters
Garment Type
LCD Display
Graphic
Inflation Pressure
(mmHg)
Inflation Hold
(seconds)
Cycle
(seconds)
Foot
130
3
30
Single-chamber,
Calf and Thigh/Calf
40
12.5
60
No Garment Connected
-
-
-
Содержание FLOWTRON UNIVERSAL
Страница 1: ...Service Manual FLOWTRON UNIVERSAL ...
Страница 2: ......
Страница 4: ...Flowtron Universal Service Manual Ammendment Record Issue 2 Page ii SER0009 Sep 2009 ...
Страница 8: ...Flowtron Universal Service Manual Issue 2 Page vi SER0009 Sep 2009 ...
Страница 22: ...Maintenance Flowtron Universal Service Manual Chap 3 Issue 2 Page 4 SER0009 Sep 2009 ...
Страница 38: ...Testing Flowtron Universal Service Manual Chap 4 Issue 2 Page 16 SER0009 Sep 2009 ...
Страница 88: ...Pump Repair Flowtron Universal Service Manual Chap 5 Issue 2 Page 50 SER0009 Sep 2009 ...
Страница 108: ...Updating the software Flowtron Universal Service Manual Chap 6 Issue 2 Page 20 SER0009 Sep 2009 ...
Страница 114: ...Parts List Flowtron Universal Service Manual Chap 8 Issue 2 Page 4 SER0009 Sep 2009 ...
Страница 116: ...SER0009_02 September 2009 ...