Flowtron Universal Service Manual
Introduction
Issue 2
Chap 1
Sep 2009
SER0009
Page 5
4.7
To start the therapy, press the green
Run
button
on the top of the pump, and keep it
depressed until the pump “beeps”. The green LEDs on the front of the pump will illuminate and
therapy will begin. Figure 5 shows the pump running, and the left calf garment inflated to 40
mmHg.
Note:
The green LEDs do not illuminate when running from the battery pack.
Figure 5 - LCD Display - Therapy Started
4.8
To stop the therapy, press the green
Run
button
a second time, and keep it depressed until
the pump “beeps”. The green LEDs on the front of the pump will extinguish and therapy will stop.
4.9
To connect only one garment to the pump:
4.9.1
When the pump is in standby, connect one garment to either hose set assembly.
4.9.2
The LCD display will confirm that only one garment is connected (Refer to Figure 6).
Note:
If a garment is disconnected while the pump is running, the system will alarm
Lo
. If the single garment option is still required, simply press the
Run
button
to reset the system and press the
Run
button
again to restart therapy.
Note:
If at any time a second garment is connected while the pump is running, the
pump will automatically change to two garment operation.
Figure 6 - LCD Display - Single Garment Connected
4.10
To connect different garments on each limb:
4.10.1
Simply connect the garments to either hose set assembly, and the correct therapy
cycle will be administered for each garment.
4.10.2
Check on the display panel to confirm the garment types connected (Refer to Figure 7).
Figure 7 - LCD Display - Different Garments Connected
Содержание FLOWTRON UNIVERSAL
Страница 1: ...Service Manual FLOWTRON UNIVERSAL ...
Страница 2: ......
Страница 4: ...Flowtron Universal Service Manual Ammendment Record Issue 2 Page ii SER0009 Sep 2009 ...
Страница 8: ...Flowtron Universal Service Manual Issue 2 Page vi SER0009 Sep 2009 ...
Страница 22: ...Maintenance Flowtron Universal Service Manual Chap 3 Issue 2 Page 4 SER0009 Sep 2009 ...
Страница 38: ...Testing Flowtron Universal Service Manual Chap 4 Issue 2 Page 16 SER0009 Sep 2009 ...
Страница 88: ...Pump Repair Flowtron Universal Service Manual Chap 5 Issue 2 Page 50 SER0009 Sep 2009 ...
Страница 108: ...Updating the software Flowtron Universal Service Manual Chap 6 Issue 2 Page 20 SER0009 Sep 2009 ...
Страница 114: ...Parts List Flowtron Universal Service Manual Chap 8 Issue 2 Page 4 SER0009 Sep 2009 ...
Страница 116: ...SER0009_02 September 2009 ...