24
Removing the Control Printed Circuit Board (PCB) Assembly
CAUTION: Electrostatic discharge can seriously damage the control PCB assembly.
Adequate earthing/grounding precautions must be taken when handling it.
24.1
Remove the hose set assemblies (Refer to Page 9, Section 8).
24.2
Remove the motor and manifold assembly from the control PCB assembly (Refer to
Page 19, Section 18).
24.3
Carefully pull the membrane cable out of the membrane connector on the control PCB assembly
(Fig 48, PCB Con 10).
24.4
Remove the four screws (Fig 48, Item 30), and lift the control PCB assembly (Fig 48, Item 10)
out of the top case assembly (Fig 48, Item 20).
Note:
Do not remove the four smaller screws (Fig 48, Item 40) which secure the LCD display to
the control PCB assembly.
25
Installing the Control Printed Circuit Board (PCB) Assembly
25.1
Put the top case assembly (Fig 48, Item 20) on a horizontal surface, with the inside surface
pointing upwards, and position the control PCB assembly (Fig 48, Item 10) just above it.
25.2
Carefully push the membrane cable into the membrane connector on the control PCB assembly
(Fig 48, PCB Con 10).
25.3
Make sure the
Run
button (Fig 48, Item 50) is installed into the cutout in the top case assembly,
and the button return spring (Fig 48, Item 60) is installed on the
Run
button.
25.4
Lower the control PCB assembly onto the four pillars in the top case assembly.
Note:
Make sure the membrane cable is positioned in the cutout in the control PCB assembly.
Note:
As the control PCB assembly is lowered onto the top case assembly, make sure the
Run
switch on the control PCB assembly stays in the centre of the button return spring on the
top case assembly (Refer to Figure 49).
25.5
Install the four screws (Fig 48, Item 30) to secure the control PCB assembly.
25.6
Install the motor and manifold assembly onto the control PCB assembly, and back into the pump
(Refer to Page 25, Section 20).
25.7
Install the hose set assemblies (Refer to Page 10, Section 9).
Содержание FLOWTRON UNIVERSAL
Страница 1: ...Service Manual FLOWTRON UNIVERSAL ...
Страница 2: ......
Страница 4: ...Flowtron Universal Service Manual Ammendment Record Issue 2 Page ii SER0009 Sep 2009 ...
Страница 8: ...Flowtron Universal Service Manual Issue 2 Page vi SER0009 Sep 2009 ...
Страница 22: ...Maintenance Flowtron Universal Service Manual Chap 3 Issue 2 Page 4 SER0009 Sep 2009 ...
Страница 38: ...Testing Flowtron Universal Service Manual Chap 4 Issue 2 Page 16 SER0009 Sep 2009 ...
Страница 88: ...Pump Repair Flowtron Universal Service Manual Chap 5 Issue 2 Page 50 SER0009 Sep 2009 ...
Страница 108: ...Updating the software Flowtron Universal Service Manual Chap 6 Issue 2 Page 20 SER0009 Sep 2009 ...
Страница 114: ...Parts List Flowtron Universal Service Manual Chap 8 Issue 2 Page 4 SER0009 Sep 2009 ...
Страница 116: ...SER0009_02 September 2009 ...