Flowtron Universal Service Manual
Testing
Issue 2
Chap 4
Sep 2009
SER0009
Page 13
Figure 30 - Inserting the Test Probe through the Rear Case
Figure 31 - Earth/Ground Test Area on Control PCB Assembly
Test Probe
Make sure the mains/power
cordset connector is pushed
securely into the pump
Mains/Power Cordset
Screw
Case IEC Cover
Note:
Insert the end of the test probe through the test hole in the rear case.
Rear Case of Pump
Approximate dimensions of test probe: Minimum length = 100 mm
Maximum diameter = 2 mm
Earth/Ground
Access Point
Control PCB Assembly
Compressor Assembly
Earth/Ground Test Pad
IEC Connector
Front Case of Pump
(inside dotted line)