Troubleshooting
Flowtron Universal Service Manual
Chap 2
Issue 2
Page 2
SER0009
Sep 2009
Table 2 - Troubleshooting Table
Fault
Possible Cause
Remedy
“LO” (Low
Pressure
Alarm)
Garment
Hose disconnected at garment.
Check the hose connections at
the garment end.
Loose fitting garment.
Rewrap snugly.
Garment leak.
Check garment and replace if
necessary.
Pump
Kinked tube to the transducer
on the control PCB.
Check tube for kinks or
obstructions.
Damaged/disconnected internal
tube.
Check internal tubes and replace
if necessary.
Damaged hose set assembly.
Check hose set assemblies, and
replace if necessary.
(Chapter 5, Page 9, Section 8)
(Chapter 5, Page 10, Section 9)
Blocked air filters.
Replace inlet filters on the rear
case and compressor assembly.
(Chapter 5, Page 7, Section 5)
(Chapter 5, Page 17, Section 17)
Motor and manifold assembly
failed.
Replace.
(Chapter 5, Page 19, Section 18)
(Chapter 5, Page 25, Section 20)
Compressor assembly failure or
low output.
Replace.
(Chapter 5, Page 45, Section 38)
(Chapter 5, Page 45, Section 39)
“HI” (High
Pressure
Alarm)
Garment
Garment hose kinked.
Check garment hose for kinks or
obstructions.
Pump
Hose set assembly kinked.
Check hose set assemblies for
kinks or obstructions.
“Pump Fault”
Pump
Control PCB faulty.
Replace.
(Chapter 5, Page 27, Section 22)
(Chapter 5, Page 29, Section 25)
Power Supply PCB faulty.
Replace.
(Chapter 5, Page 31, Section 26)
(Chapter 5, Page 31, Section 27)
“Pump Fault -
RFID”
Pump
Hose set electrical connections
are not fitted correctly to pump.
Check hose set assemblies, and
replace if necessary.
(Chapter 5, Page 9, Section 8)
(Chapter 5, Page 10, Section 9)
“Pump Fault -
Step”
Pump
Motor and manifold assembly
failed.
Replace.
(Chapter 5, Page 19, Section 18)
(Chapter 5, Page 25, Section 20)
Содержание FLOWTRON UNIVERSAL
Страница 1: ...Service Manual FLOWTRON UNIVERSAL ...
Страница 2: ......
Страница 4: ...Flowtron Universal Service Manual Ammendment Record Issue 2 Page ii SER0009 Sep 2009 ...
Страница 8: ...Flowtron Universal Service Manual Issue 2 Page vi SER0009 Sep 2009 ...
Страница 22: ...Maintenance Flowtron Universal Service Manual Chap 3 Issue 2 Page 4 SER0009 Sep 2009 ...
Страница 38: ...Testing Flowtron Universal Service Manual Chap 4 Issue 2 Page 16 SER0009 Sep 2009 ...
Страница 88: ...Pump Repair Flowtron Universal Service Manual Chap 5 Issue 2 Page 50 SER0009 Sep 2009 ...
Страница 108: ...Updating the software Flowtron Universal Service Manual Chap 6 Issue 2 Page 20 SER0009 Sep 2009 ...
Страница 114: ...Parts List Flowtron Universal Service Manual Chap 8 Issue 2 Page 4 SER0009 Sep 2009 ...
Страница 116: ...SER0009_02 September 2009 ...