
Flowtron Universal Service Manual
Pump Repair
Issue 2
Chap 5
Sep 2009
SER0009
Page 43
36
Removing the Compressor Assembly and Mounting Bracket
36.1
Remove the rear case from the pump (Refer to Page 13, Section 10).
36.2
Lay the front case down with the inside surface pointing upwards.
36.3
Disconnect the large diameter tube (Fig 58, Item 50) from the compressor assembly.
36.4
Disconnect the two compressor cable connectors (Fig 58, PCB Con 6 and 7) from the power
supply PCB assembly (Fig 58, Item 40).
36.5
Remove the two screws (Fig 58, Item 30), and lift the compressor assembly and mounting
bracket (Fig 58, Items 10 and 20) out of the front case.
37
Installing the Compressor Assembly and Mounting Bracket
37.1
Before installing the compressor assembly and mounting bracket (Fig 58, Items 10 and 20),
check the following parts are installed correctly on it (Refer to Figure 61):
37.1.1
The air inlet filter into the air inlet filter holder.
37.1.2
The eight push-on cable connectors, from the power supply PCB assembly
(Fig 58, PCB Con 6 and 7).
37.2
Put the compressor assembly and mounting bracket into the front case.
Note:
Make sure the two locating holes at the outer ends of the mounting bracket are installed
over the corresponding location spigots “A” and “B” in the front case (Refer to Figure 58).
Note:
Make sure the wires from the power supply PCB assembly to the battery pack connector
do not lie on top of the circular compressor support rubbers (Refer to Figure 50).
37.3
Install the two screws (Fig 58, Item 30) through the mounting bracket (Fig 58, Item 20) and into
the front case.
Note:
One hole in the left side of the mounting bracket is not used (Refer to Figure 58).
37.4
Connect the large diameter tube (Fig 58, Item 50) onto the compressor assembly.
Note:
Make sure the other end of the tube is still connected to the correct inlet tube on the
motor and manifold assembly (Figure 41).
37.5
Connect the two compressor cable connectors (Fig 58, PCB Con 6 and 7) onto the power
supply PCB assembly (Fig 58, Item 40).
37.6
Install the rear case onto the pump (Refer to Page 13, Section 11).
Содержание FLOWTRON UNIVERSAL
Страница 1: ...Service Manual FLOWTRON UNIVERSAL ...
Страница 2: ......
Страница 4: ...Flowtron Universal Service Manual Ammendment Record Issue 2 Page ii SER0009 Sep 2009 ...
Страница 8: ...Flowtron Universal Service Manual Issue 2 Page vi SER0009 Sep 2009 ...
Страница 22: ...Maintenance Flowtron Universal Service Manual Chap 3 Issue 2 Page 4 SER0009 Sep 2009 ...
Страница 38: ...Testing Flowtron Universal Service Manual Chap 4 Issue 2 Page 16 SER0009 Sep 2009 ...
Страница 88: ...Pump Repair Flowtron Universal Service Manual Chap 5 Issue 2 Page 50 SER0009 Sep 2009 ...
Страница 108: ...Updating the software Flowtron Universal Service Manual Chap 6 Issue 2 Page 20 SER0009 Sep 2009 ...
Страница 114: ...Parts List Flowtron Universal Service Manual Chap 8 Issue 2 Page 4 SER0009 Sep 2009 ...
Страница 116: ...SER0009_02 September 2009 ...