Flowtron Universal Service Manual
Pump Repair
Issue 2
Chap 5
Sep 2009
SER0009
Page 11
Figure 37 - Installing the Hose Set Connectors
9.6
When the elbow retaining cover is installed into the pump case cutout, turn the elbow retaining
cover approximately 20 degrees clockwise, to the “secured” position (Refer to Figure 38).
Note:
To turn the elbow retaining cover more easily, use the tube and elbow connector to
provide extra leverage. Turn the tube and elbow connector counterclockwise until the
elbow connector touches the screw cover, and then continue turning them to turn the
elbow retaining cover (Refer to Figure 38).
9.7
When the elbow retaining cover is in the correct position, install the screw (Fig 34, Item 20) and
tighten it to secure the hose set assembly to the pump.
9.8
Carefully press down the screw cover until it clicks into position (Refer to Figure 35).
Figure 38 - Securing the Hose Set Connector
Elbow Retaining Cover
Lug
Hose Set Cable Connector
Control Board Connector
Pneumatic Outlet
Pump Case
Case Cutout for Elbow Retaining Cover
Elbow Connector
Insert excess hose set
cable this side of control
board connector to prevent
damage by the lugs on
elbow retaining cover
Elbow Retaining Cover
in Secured Position
Elbow Retaining Cover
in Free Position
20 degrees
Tube
Elbow Connector
Turn Tube and
Elbow Connector
clockwise
Elbow
Retaining
Cover
Screw Cover
Содержание FLOWTRON UNIVERSAL
Страница 1: ...Service Manual FLOWTRON UNIVERSAL ...
Страница 2: ......
Страница 4: ...Flowtron Universal Service Manual Ammendment Record Issue 2 Page ii SER0009 Sep 2009 ...
Страница 8: ...Flowtron Universal Service Manual Issue 2 Page vi SER0009 Sep 2009 ...
Страница 22: ...Maintenance Flowtron Universal Service Manual Chap 3 Issue 2 Page 4 SER0009 Sep 2009 ...
Страница 38: ...Testing Flowtron Universal Service Manual Chap 4 Issue 2 Page 16 SER0009 Sep 2009 ...
Страница 88: ...Pump Repair Flowtron Universal Service Manual Chap 5 Issue 2 Page 50 SER0009 Sep 2009 ...
Страница 108: ...Updating the software Flowtron Universal Service Manual Chap 6 Issue 2 Page 20 SER0009 Sep 2009 ...
Страница 114: ...Parts List Flowtron Universal Service Manual Chap 8 Issue 2 Page 4 SER0009 Sep 2009 ...
Страница 116: ...SER0009_02 September 2009 ...