46
Every Month
Test the thermostatic mixer
Turn the temperature control back and forth over the entire
temperature range with water on. This is particularly
important in areas with hard water.
Make sure the temperature is changing. See Fig. 1
Check/clean filters for incoming water:
A shut-off valve with a filter is mounted on each of the two
water supply hoses.
To access these:
1 Make sure the
Malibu/Sovereign
bathtub is in its highest
position.
2 The shut-off valve can be reached under the
Malibu/Sovereign
bathtub.
The filter in both valves should be rinsed as follows:
1 Close the valves (
A
). The handles (red for hot water and
blue for cold) should be in a cross position. See Fig. 2
2 Remove the cap (
B
) with a wrench.
3 Pull out the filter (
C
) and clean it carefully in water.
4 Reassemble (
B
) and (
C
).
5 Open the valves. (
A
)
WARNING
Pressurised system and risk of scalding. Make sure
that the valve (A) is closed before removing the cap,
this is especially important for the hot water valve.
See Fig. 2
Fig. 1
Fig. 2
A
B
C
Содержание Malibu