![background image](http://html1.mh-extra.com/html/arjo/malibu/malibu_instructions-for-use-manual_2972247036.webp)
36
Microbiological Formation Prevention
In order to enhance the on-site curative corrective
actions (chlorine or thermal shock), which do not
guarantee a reduction of the contamination long
term, make sure to follow the instructions below.
Installation precautions
In order to limit the growth of bacterial flora, it is
necessary to take action at four levels:
•
Avoid stagnating water and make sure proper
water circulation;
•
Take precautions against the formation of scale
and corrosion based on the water quality;
•
Maintain a high water temperature all the way
from the hot water dispenser to the different
distribution pipes;
•
Mix hot and cold water as close as possible to the
point of use.
Use
•
Make sure that the water circulates in the
Malibu/
Sovereign
bath and the shower on a daily basis
even if the bathtub is not used; and particular to
make sure to remove any water that may be left
behind in the hose.
•
Let the water flow approximately 5 minutes
before the first
Malibu/Sovereign
bath of the day.
•
Clean and disinfect the
Malibu/Sovereign
bathtub
according to the
IFU
before the first bath of the
day and after the bath of each patient.
Do not forget to contact Arjo for the annual