AZI Customer Service 800-528-7411 or 602-470-1414 Page
9
of
21
CAL 1 [Auto-Calibration] - Using an external calibration weight
Perform auto-calibration every time you move your balance. It is extremely important to
use high quality weights and verify the balance has been warmed up for at least one (1)
hour prior to calibration.
Procedure
Display Reading
Remove any containers or weights so
that only the stainless steel pan is in
place. Press the ZERO switch.
Zeros are displayed.
Press the FUNCTION switch.
Display cycles repeatedly through PCS, HI-
OK-LO, CAL1, and CAL2.
Press the MODE switch when CAL 1
appears.
CAL 1 and a flashing 0 are displayed.
Wait 10 seconds then press the ZERO
switch.
The display stops flashing. In approximately
10 seconds a weight value or two alternating
weight values flash on the display.
Place the displayed weight or one of
the alternating weights on the pan.
The display stops flashing and the selected
weight are displayed.
Wait 10 seconds then press the ZERO
switch.
The weight continues to be displayed while
the calibration takes place (approximately 10
seconds). The display blanks, then displays
"OK" and the calibration weight.
Remove the calibration weight.
The display returns to zeros and the normal
weighing mode.