AZI Customer Service 800-528-7411 or 602-470-1414 Page
10
of
21
Capacity Tracker
The capacity tracker provides a graphic display of the used and unused portions of the
weighing range. Each segment represents 5% of the balance’s capacity and is
illuminated as that portion of the capacity is used. The larger segments, representing
¼, ½, ¾ and full load, flash as they are approached and then become solid again as
they are exceeded.
Selectable Vibration Filters and Stability Indicator
All balances are equipped with three vibration filters and a stability indicator to adapt to
varying weighing conditions. The balance is delivered set in the high (Fil HI) mode,
which is the most aggressive mode. Less filtering (Fil nor or Fil Lo) can be selected for
quieter conditions and faster response.
The stability indicato
r “OK” is illuminated when the balance has determined that all
readings are within the filter parameters the user has selected. If the weight readings
are outside the acceptable limit, the stability indicator will not light. The stability
indicator can be disabled if desired.
To enter the vibration filter selection mode, proceed as follows:
Procedure
Display Reading
Turn balance OFF using the ON/OFF
switch.
Display blanks.
Hold the MODE switch down while turning
balance on using the ON/OFF switch.
OK and the current filer, Fil Nor, balance
Fil Lo or Fil HI is displayed.
Press the SEND switch until the desired
vibration filter is displayed. Then press
the ZERO switch.
OK SAVEd is displayed followed by the
current stability indicator setting, Si on or
Si oFF.
Press the SEND switch until the desired
stability indicator setting is displayed and
then press the ZERO switch.
OK SAVEd is displayed followed by the
vibration filter setting chosen in step 3.
Press the ON/OFF switch.
The display blanks.
Press the ON/OFF switch.
The balance returns to normal operation
with the new vibration filter and stability
indicator settings in place.