AZI Customer Service 800-528-7411 or 602-470-1414 Page
12
of
21
RS-232 Front Panel Configuration
The RS-232 Configuration Mode allows changes to the baud rate, front panel control,
software protocol, parity and auto send parameters. The balance is preconfigured at
Arizona Instrument to communicate with a Vapor Pro.
Any changes to the recommended configuration will affect
communication with the Vapor Pro.
It is not necessary to follow the entire configuration procedure if only one parameter is
to be changed. Follow steps 1 and 2, and then use the MODE switch to cycle through
the different parameters. When the parameter to be changed is displayed, toggle the
setting using the SEND switch until the desired setting is displayed. Save the setting
using the ZERO switch, and the balance will co
nfirm by displaying “SAVEd”. Then,
turn the balance off and then on again with the ON/OFF switch to return to normal
operating mode.
To enter RS-232 configuration mode, proceed as follows:
Procedure
Display Reading
Turn balance off using the ON/OFF
switch
Display blanks
Hold the SEND switch down while
turning balance on using ON/OFF switch
The current baud rate is displayed
Press the SEND switch until br 9600 is
displayed.
br 9600
Press the ZERO switch to save br 9600
as the BAUD RATE.
SAVEd
is displayed followed by
br 9600
.
Press the MODE switch to move to the
next parameter: LOCAL CONTROL
The current LOCAL CONTROL setting, LC
on or LC oFF is displayed.
Press the SEND until LC oFF is
displayed.
LC oFF
Press the ZERO switch to save LC oFF
as the LOCAL CONTROL setting.
SAVEd
is displayed followed by
LC oFF
.
Press the MODE switch to move to the
next parameter: SOFTWARE
PROTOCOL
The current SOFTWARE PROTOCOL
setting, Pr oFF or Pr on is displayed.
Press the SEND switch until Pr oFF is
displayed.
Pr oFF
Press the ZERO switch to save Pr oFF
as the SOFTWARE PROTOCOL setting.
SAVEd
is displayed followed by
Pr oFF
.
Press the MODE switch to move to the
next parameter: PARITY
The current parity setting, PA EVEn, PA
odd or PA none is displayed.
Press the SEND switch until PA none is
displayed.
PA none