AZI Customer Service 800-528-7411 or 602-470-1414 Page
20
of
21
Appendix B - Weight Checking
The following procedure checks the weight of individual parts against user-
programmed upper and lower weight limits.
Procedure
Display Reading
Press the FUNCTION switch.
Display cycles repeatedly through PCS, HI-OK-
LO, CAL1, and CAL2
Press the MODE switch when HI-OK-LO
appears on the display.
HI, LO and O flash.
Press the ZERO switch.
HI flashes and zeros are displayed.
Place weight on pan that represents the
upper limit of the acceptable weight span.
HI flashes and weight is displayed.
Press the ZERO switch.
LO flashes and upper limit weight is displayed.
Remove upper limit weight and place
weight on the pan that represents the
lower limit.
LO flashes and weight is displayed.
Press the ZERO switch.
HI-OK-LO briefly appears then OK and the low
limit weight are displayed.
Remove low limit weight.
LO appears and zeros are displayed.
The balance is now ready to check the weight of parts. The balance displays OK
along with the weight of the part if the part is within the selected limits. If the part
is too heavy, HI and the weight are displayed. If the part is too light, LO and the
weight are displayed. The limits are stored until the balance is turned off. To exit
weight checking, press the MODE switch and the balance will return to normal
operation.