FITOCLIMA 2500 HP
Mod. MI 271 EN / R00 23-10-2015
-67-
•
Lamps thermostat –
Lamps thermal circuit breaker failure;
•
Low suction temperature –
Compressor suction temperature below factory default
setting (typically -20ºC). The alarm stops when the temperature raise 5°C compared
to the factory set value;
•
Ventilation 2 power fail –
Ventilador 2 power supply failure;
•
Bypass damper position –
Bypass damper position failure;
•
Air damper position –
Air damper position failure;
•
CO2 sensor fail –
CO2 sensor failure;
•
Evaporator out sensor fail –
Evaporator output sensor failure;
•
Evaporator 2 sensor fail –
Evaporator 2 sensor failure;
•
Evaporator 2 out sensor fail –
Evaporator 2 output sensor failure;
•
Lamps temperature sensor fail –
Lamps temperature sensor failure;
•
Mobile temperature sensor fail –
Mobile temperature sensor failure;
•
Radiation sensor fail –
Radiometer sensor failure;
•
High radiation –
Radiation intensity above the maximum set value, during the time
set for alarm delay;
•
Low radiation –
Radiation intensity below the maximum set value, during the time
set for alarm delay;
•
Radiation band –
Radiation intensity above the value obtained by adding the
setpoint with the band, or below the value obtained by subtracting the setpoint with
the band, during the time set for alarm delay;
•
Compressor internal fail -
Compressor thermal circuit breaker failure;
•
Igniters thermostat -
Igniters thermal circuit breaker failure;
•
Cond. water temp. sensor fail –
Refrigeration water temperature sensor failure;
•
Cond. water press. sensor fail –
Refrigeration water pressure sensor failure;
•
Cond. water high temperature –
Refrigeration water temperature above factory
default setting (typically 26ºC). The alarm stops when the temperature drops 1ºC
compared to the factory set value;
Содержание FITOCLIMA 2500 HP
Страница 1: ...USER S MANUAL READ AND KEEP THIS INSTRUCTIONS ORIGINAL MANUAL EN ...
Страница 3: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 3 ...
Страница 13: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 13 Top view ...
Страница 14: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 14 Side view ...
Страница 15: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 15 3D view ...
Страница 20: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 20 Humidifiers Allows humidifying air inside the chamber ...
Страница 39: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 39 4 4 Set values configuration ...