FITOCLIMA 2500 HP
Mod. MI 271 EN / R00 23-10-2015
-64-
•
Humidifier water fail –
Lack of water in the humidifier for more than 20 minutes;
•
Refrigeration fail –
Failure in the refrigeration system;
•
Humidifier temperature –
Humidifier temperature above 50°C. The alarm stops
when the temperature of the humidifier register a value below 49ºC;
•
Evaporator temperature –
Evaporator temperature above 70°C. The alarm stops
when the temperature of the evaporator to register a value below 69ºC;
•
Heater temperature –
Heater temperature above 70ºC. The alarm stops when the
temperature of the resistance register a value below 65ºC;
•
Condensing temperature –
Condenser Temperature above factory default setting
(typically 60ºC). It is necessary to clean the condenser. The alarm stops when the
temperature drops 1°C compared to the factory set value;
•
Temperature sensor fail –
Air Temperature sensor failure;
•
Humidity sensor fail –
Relative Humidity sensor failure;
•
Humidifier sensor fail –
Humidifier temperature sensor failure;
•
Evaporator sensor fail –
Evaporator temperature sensor failure;
•
Heater sensor fail –
Heater temperature sensor failure;
•
Condensing sensor fail –
Condenser temperature sensor failure;
•
UC2 Communication fail –
Very serious error that indicates failure in
communications between electronic circuit boards. The thermal safety of the
equipment is compromised in this situation;
•
Air safety temperature sensor fail –
Security temperature sensor failure;
•
Temperature sensor 2 fail –
Secondary
Air Temperature sensor failure (T2);
•
Humidity sensor 2 fail –
Secondary
Relative Humidity sensor failure (H2);
•
Alarm test –
Whenever it is made a routine test in the alarm system;
•
Discharge pressure sensor fail –
Compressor discharge pressure sensor failure;
•
Discharge temperature sensor fail –
Compressor discharge temperature sensor
failure;
Содержание FITOCLIMA 2500 HP
Страница 1: ...USER S MANUAL READ AND KEEP THIS INSTRUCTIONS ORIGINAL MANUAL EN ...
Страница 3: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 3 ...
Страница 13: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 13 Top view ...
Страница 14: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 14 Side view ...
Страница 15: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 15 3D view ...
Страница 20: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 20 Humidifiers Allows humidifying air inside the chamber ...
Страница 39: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 39 4 4 Set values configuration ...