FITOCLIMA 2500 HP
Mod. MI 271 EN / R00 23-10-2015
-27-
6)
If the equipment has a water-cooled condenser, booth inlet and outlet must be connected
to a cooling water circuit,
The technical characteristics of the circuit should be as follows:
•
Water flow - up to 1000 liters / hour;
•
Intake pressure – 3 to 6 bar;
•
Inlet and outlet water section - binding pipe 3/4”;
•
Differential pressure between input and output: >2,5 bar;
•
Maximum water input temperature: +26 ºC (+30 ºC output);
•
Recommended water input temperature: +18 °C.
7)
Connect the output of sewage to the sewage of the building. This should be at ground
level and be at least Ø 50 mm. The flow is by gravity as such, from the output of the
chamber to the sewage system, and the connection must have a slope of at least 15º
downward.
8)
Connect the water inlet for irrigation system and service tap, if available.