FITOCLIMA 2500 HP
Mod. MI 271 EN / R00 23-10-2015
-48-
4.9
Program evolution
The program’s evolution is defined by Ramp ON
/
Ramp OFF:
•
If Ramp OFF each segment defines a Set-Point during the segment time.
•
If Ramp ON, evolution will be in ramp during the segment time.
Relative Humidity will evolve in ramp except when it comes from 0 (OFF) or when it goes
to OFF, that is., if in a certain segment the Relative Humidity Set-Point is 0 (OFF), even if
Ramp ON the Relative Humidity will go immediately OFF, if in a previous Relative
Humidity segment was OFF and the segment taking place is, for example 50%rH, the
Relative Humidity immediately starts controlling to 50%rH. Radiation intensity and CO2
level has similar evolution to Humidity..
Example of a program:
Segment 0:
T=15ºC, H=50%rH, 2 min
Segment 1:
T=30ºC, H=0%rH, 5 min
Segment 2:
T=30ºC, H=50%rH, 3 min
Segment 3:
T=15ºC, H=80%rH, 5 min
Segment 4:
T=15ºC, H=80%rH, 10 min
Segment 5:
T=30ºC, H=50%rH, 3 min
Segment 6:
T=30ºC, H=50%rH, 0 min
(
PG-END
)
Ramp ON
Ramp OFF
Содержание FITOCLIMA 2500 HP
Страница 1: ...USER S MANUAL READ AND KEEP THIS INSTRUCTIONS ORIGINAL MANUAL EN ...
Страница 3: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 3 ...
Страница 13: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 13 Top view ...
Страница 14: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 14 Side view ...
Страница 15: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 15 3D view ...
Страница 20: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 20 Humidifiers Allows humidifying air inside the chamber ...
Страница 39: ...FITOCLIMA 2500 HP Mod MI 271 EN R00 23 10 2015 39 4 4 Set values configuration ...