MAINTENANCE
- At the end of each treatment, let clean water flow inside the pipe.
- In case of malfunction, and at any rate at the end of every season, clean the flowmeter pipe with a special detergent.
- Do not use metal or abrasive objects to clean the pipe.
- Do not use solvents or fuel to clean the case outer surface.
8
ERRORS, TROUBLESHOOTING
Trouble
"Displayed error
(for version with Display,
only)"
LED status
Number
of b
links
Cause
Remedy
No flowrate
measurement
---
Off
---
No power supply.
Check the connections and the harnesses.
---
---
Replace the cable. If the problem persists,
contact your nearest Service Center.
---
---
Voltage is below the minimum permitted
value.
Check the power supply voltage and restore the
correct value.
---
Green steady on
---
Flowrate is below the minimum
measurable value
Check that the liquid passes through the
flowmeter and that the flowrate is above the
minimum measurable value.
---
Blue steady on
---
CanBus
The canbus address assigned to the
sensor is wrong.
Repeat the address assignment procedure.
---
---
The CanBus network harness is not
suitable.
Make sure that the harness structure complies
with the CanBus 2.0 specifications.
---
---
Check for the presence and conditions of
terminating resistors.
---
---
Fr
equenc
y
The connection is wrong.
Check connections.
---
---
The type of input of the connected
reading device is not compatible with
the Orion X output stage
Make sure that the technical specifications of the
two devices are compatible.
---
Violet steady on
---
Orion X does not detect the presence of
liquid inside pipe.
Make sure that there is liquid inside the
flowmeter.
---
---
Make sure that liquid conductivity is higher than
the one stated in the technical specifications.
---
White steady on
---
Orion X is cleaning the heads and
cannot measure the flowrate.
Turn the flowmeter off and on again to restart the
measurement.
Hw errors detected
Red blinking
2
Hardware error.
Contact an ARAG authorized Service Center.
Wrong calibration
3
Corrupted calibration data.
Contact an ARAG authorized Service Center.
Voltage out of limit
4
Power supply voltage too high or too low.
Check the power supply voltage and restore the
correct value.
Unstable reading
5
Unstable reading.
Check that the flow is normal and make sure
to have complied with the hydraulic connection
instructions (position, pipe dimensions, straight
sections before and after the flowmeter, etc.).
High temperature!
6
Too high temperature has been detected.
Check ambient temperature and, if outside the
specified limits, turn off the device and, before
turning it on again, wait for the temperature
conditions to fall within the specified ones.
Check parameters
7
Setting data are corrupted and the
default ones have been restored.
Contact an ARAG authorized Service Center
to check data integrity and to restore correct
operation.
Check that all settings are as required, validate
data by accessing the "New notifications" menu
and press OK (the operation can be performed
only by skilled technical staff).
Unstable
measurement
---
Blue steady on
---
Wrong installation.
Check that the flow is normal and make sure
to have complied with the hydraulic connection
instructions (position, pipe dimensions, straight
sections before and after the flowmeter, etc.).
The measured
flowrate does not
correspond to the
actual one
---
---
The measuring pipe shows chemical
deposits.
Fill the pipe with clean water and activate the
electrode cleaning procedure for at least 8 hours.
---
---
Clean with a soft cloth and denatured ethyl
alcohol.
The display
shows an error
Stop flow!
Blue steady on
---
In filling mode, the device sent the pump
stop command, but the sensor continues
measuring a flowrate.
Make sure that the flow is interrupted by the
pump of by the connected valve.
Low flowrate!
---
Flowrate below the minimum set
threshold.
Check that the actual flowrate is within the set
range.
Check the set values.
High flowrate!
---
Flowrate above the maximum set
threshold.
"WARNING!
WRONG V.x.x.x!
Upgrade OrionX to the
V x.x.x Firmware Version"
LED ON
regardless of the color
---
Obsolete firmware version.
Contact an ARAG authorized Service Center to
check whether the firmware has to be updated.
Check sensor communic.!
---
The display does not detect the sensor.
Restart the flowmeter, if the problem persists,
contact an ARAG authorized Service Center.