SECTION 7: CONNECTORS, APPLICATIONS & PROTECTION FEATURES
Entire Contents Copyright
2014 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
42L Series Modular DC Load Operation Manual
Page 55 of 88
KEY# Description
2.
OPP test function parameter setting key. There are four parameters for the OPP test
function. These parameters are: PSTAR, PSTEP, PSTOP and Vth. Press the “OPP” key again
to set the OPP test parameter PSTAR (start power point). Press the “OPP” key again to
proceed to the next parameter in order of PSTEP, PSTOP, Vth and Disable. Press any other
key to exit and save the OPP setting.
The OPP test parameter descriptions are as follows:
a.
PSTAR: setting the start power, The LCD display shows “OPP”, “PSTAR” and 0.00W
(or initial setup value) from top to bottom 5 digits LCD display, the setting range is
0.00W to the full scale of the CP mode specification. The setting is by rotating the
dial.
b.
STEP: setting the increment step power, The LCD display shows “OPP”, “PSTEP”
and 0.00W (or initial setup value) from top to bottom 5 digits LCD display, the
setting range is 0.00W to the full scale of the CP mode specification. The setting is
by rotating the dial.