SECTION 7: CONNECTORS, APPLICATIONS & PROTECTION FEATURES
Entire Contents Copyright
2014 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
42L Series Modular DC Load Operation Manual
Page 48 of 88
KEY# Description
Setting the Vsense mode:
Press the CONFIG key. The middle 5 digit LCD display will show "SENSE".
The lower 5 digit LCD display will show "AUTO" or "ON". Use the rotary dial to toggle the setting.
Setting the LOAD ON voltage levels:
Press the CONFIG key twice. The middle 5 digit LCD display will
show "LDon". The lower 5 digit LCD display will show the set value in Volts. The Load ON Voltage set
range for all 42L modules is:
42L08xx
0V to 25V in 0.1V steps
Adjust by rotating the dial. If the input voltage to the load is greater than the Load ON voltage
setting, the load starts to sink current.
Setting the LOAD OFF voltage levels:
Press the CONFIG key till the LOAD OFF screen appears. The
middle 5 digit LCD display will show "LDoff". The lower 5 digit LCD display will show the set value in
Volts. The Load OFF Voltage set range varies by 41L model:
41L0615 ,41L0630, 42L0860
0V to 24.9V in 0.1V steps
41L2512
0V to 49.8V in 0.2V steps
41L5012
0V to 99.5V in 0.5V steps
Adjust by rotating the dial. If the input voltage to the load is lower than the Load OFF voltage
setting, the load turns off.
Setting the Load polarity:
Press the CONFIG key till the POLARITY screen appears. The middle 5 digit
LCD display will show "POLAR". The lower the 5 digit LCD display will show “+ LOAD" or "-LOAD".
Use the dial to toggle between "+ LOAD" or "-LOAD" settings.