SECTION 7: CONNECTORS, APPLICATIONS & PROTECTION FEATURES
Entire Contents Copyright
2014 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
42L Series Modular DC Load Operation Manual
Page 50 of 88
KEY# Description
b.
V-Hi: Short test voltage check upper limitation setting. The LCD displays show
“SHORT”, “V-Hi” and 0.00V (or initial setup value) from top to bottom. The V-Hi setting
range is from 0.00 to 60.00V in 0.01V steps and can be adjusted by turning the rotary
knob.
c.
V-Lo: Short test voltage check lower limitation setting. The LCD displays show
“SHORT”, “V-Lo” and 0.00V (or initial setup value) from top to bottom. The V-Lo
setting range is from 0.00 to 60.00V in 0.01V steps and can be adjusted by turning the
rotary knob.
Note:
The V-Hi and V-Lo parameters of the SHORT test are different from the V-Hi and V-Lo settings
of the LIMIT function.