Working Standard
User’s Guide
www.appliedp.com
WS-UGFW619-EN
79
SYST:ENER:ALG {<algorithm>}
Setting of the method of summing energy. Parameter can be 0 = 4Quadrant, 1 = Algo1 (Net Result), 2
= Algo2 (Positive Aggregate), 3 = Algo3 (Both Sum), 4 = Algo4 (Anti-fraud).
SYST:ENER:SENS?
Request for the setting of energy direction during summing energy. Only for Algorithm different from
4Quadrant (see
SYST:ENER:ALG
command). Possible answers are 0 = Import, 1 = Export.
SYST:ENER:SENS {<sense>}
Setting of the energy direction during summing energy. Only for Algorithm different from 4Quadrant
(see
SYST:ENER:ALG
command). Parameter can be 0 = Import, 1 = Export.
SYST:ENER:IMP?
Request for the setting of the multiplication constant on the device impulse output. Reply is real number
with
the value of the set multiplication constant (e.g. “+2.02”). The unit of this constant is 1/Ws
or 1/VAs .
SYST:ENER:IMP {<factor>}
Setting of the multiplication constant of the device impulse output which generates the impulse signal
proportional to the measured energy. Parameter <factor> is the multiplication constant of the impulse
output in the unit of 1/Ws or 1/VAs.
SYST:ENER:IMP:K?
Request for the setting of the multiplication constant on the device’s impulse output. Reply is real
number with the value of the set multiplication constant (e.g. “+2.02”). The unit of this
constant is 1/kWh or 1/kVAh.
SYST:ENER:IMP:K {<factor>}
Setting of the multiplication constant of the device’s impulse output which generates the impulse signal
proportional to the measured energy. Parameter <factor> is the multiplication constant of the impulse
output in the unit of 1/kWh or 1/kVAh.
SYST:ENER:IMP:STAT?
Request for the state of the device impulse output. Reply is number:
0 … state is OFF
1 … state is ACTIVE POWER
2 … state is REACTIVE POWER
3 … state is FIXED
4 … state is APPARENT POWER
SYST:ENER:IMP:STAT {0|1|2|3|4} | {OFF|ACTIVE|REACTIVE|FIXED|APPARENT}
Setting of the device impulse output state. Parameter is state code (0 to 4 listed in previous command)
or state name.
SYST:ENER:IMP:FIX?
Request for the constant frequency value (which is generated if the impulse output is in the state
FIXED). Reply is the text string containing a real number with the value of this frequency (e.g.
“+1.03”). This value is in a Hz unit.
SYST:ENER:IMP:FIX {<freq>}
Setting of the constant frequency value (which is generated if the impulse output is in the state FIXED).
Parameter <freq> is the frequency value in Hz unit.