Working Standard
User’s Guide
WS-UGFW619-EN
www.appliedp.com
42
4.7.1 Test Result Details
Measured samples can be seen in Test Result Details page which can be activated by
F3
from Meter
Error Test page. This screen shows all test parameters and all measured samples with pulses or
energies against time used for evaluation of these samples.
Samples screen
* P66
Content:
SN
.................... serial number of meter
Error
................. resulting error
Deviation
......... standard deviation of last completed test
Impulses
.......... number of (required) impulses
Energy
............. energy being measured during last
completed test
Samples:
No
..................... Sequence number
Error [%]
........... Measured sample (error in %)
The last column toggles between (
F5
):
Exp. / Meas imp.
...... Expected impulses count / Measured
impulses count or
or
E[kWh] / Time [s]
.....
measured energy over time (unit
changes with Measurement Settings)
*1
*
1
Up to version 5.30 (including) of FW the quantity E[xxx] / Time [s]
with different meaning „calculated energy over time
with respect to Energy and Error [%]”.
Control keys:
F1, ESC
.. Exit (return to previous page)
F3
............. Enables to store recent test results to new meter (Add meter screen will appear)
F4
............. Saves recent test results to database for currently selected meter
F5
............. Toggles between
Exp. / Meas imp.
and
E[kWh] / Time [s]
view in third column