Working Standard
User’s Guide
www.appliedp.com
WS-UGFW619-EN
91
MEAS:ENER:REAC:CHAN? {<channel (0|1|2)>}
Request for reactive energy values in vars in particular channel. Parameter channel can be 0=L1, 1=L2,
2=L3.
MEAS:ENER:REAC:CHAN:PRIM? {<channel (0|1|2)>}
Request for primary reactive energy values in vars in particular channel. Parameter channel can be
0=L1, 1=L2, 2=L3.
MEAS:ENER:REAC:CHAN:K? {<channel (0|1|2)>}
Request for reactive energy values in kvarh in particular channel. Parameter channel can be 0=L1,
1=L2, 2=L3.
MEAS:ENER:REAC:CHAN:PRIM:K? {<channel (0|1|2)>}
Request for primary reactive energy values in kvarh in particular channel. Parameter channel can be
0=L1, 1=L2, 2=L3.
MEAS:ENER:APP?
Request for apparent energy values in VAs.
MEAS:ENER:APP:PRIM?
Request for primary apparent energy values in VAs.
MEAS:ENER:APP:K?
Request for apparent energy values in kVAh.
MEAS:ENER:APP:PRIM:K?
Request for primary apparent energy values in kVAh.
MEAS:ENER:APP:CHAN? {<channel (0|1|2)>}
Request for apparent energy values in VAs in particular channel. Parameter channel can be 0=L1,
1=L2, 2=L3.
MEAS:ENER:APP:CHAN:PRIM? {<channel (0|1|2)>}
Request for primary apparent energy values in VAs in particular channel. Parameter channel can be
0=L1, 1=L2, 2=L3.
MEAS:ENER:APP:CHAN:K? {<channel (0|1|2)>}
Request for apparent energy values in kVAh in particular channel. Parameter channel can be 0=L1,
1=L2, 2=L3.
MEAS:ENER:APP:CHAN:PRIM:K? {<channel (0|1|2)>}
Request for primary apparent energy values in kVAh in particular channel. Parameter channel can be
0=L1, 1=L2, 2=L3.
MEAS:SIGN? {<phase (0|1|2)>},{<unit (0|1)>} ,{<harmonic (0-31)>}
Request for signal’s harmonic content.
Parameters: phase (0=L1, 1=L2, 2=L3), unit (0=U, 1=I) and harmonic number.
Answer: amplitude and phase (two comma separated real numbers).
Examples:
MEASure:SIGNal? 0,0,2
… reading of content of 2
nd
harmonic of U
L1
MEASure:SIGNal? 2,1,5
… reading of content of 5
th
harmonic of I
L3
1.234567E-01, 9.01
… answer for MEASure:SIGNal? Command
MEAS:SIGN:SAMP? {<phase (0|1|2)>},{< unit (0|1)>}
Содержание WS 2120