WS-UGFW619-EN
73
4.14 Test Report
Test report of selected meter can be printed on
connected portable printer (which supports used
protocol) using
F2
in Database of Tests. Printer
is
not included in default delivery but it is possible
to
order
one
from
Working
Standard’s
manufacturer.
Except the printer, such report might be sent to
other device via serial bus. This device, for
instance a PC, has to be equipped with standard
serial bus.
Test report contains:
-
date of measurement
-
parameters of tested meter
-
results of all tests (load points) of
selected meter
Example of Test Report
HEADER
DATE: 15.02.2009
SNUM: 1315050166
TYPE: ET 420
YMAN: 1998
VOLTAGE: 3 x 230/400V
CURRENT: 10-40A
CONSTANT: 375r/kWh
CLASS: 2 %
CUSTOMER: Applied Precision
ADDRES: Stavitelska 1, Bratislava
NOTE1: Note 1
NOTE2: Note 2
NOTE3: Note 3
NOTE4: Note 4
LOADPOINT 1
U[V] 230.02 229.48 229.92
I[A] 120.09 120.20 120.01
P[W] 13855.27 13855.27 13855.27
Q[var] 23873.34 23816.44 23874.72
S[VA] 27599.07 27606.00 27581.49
PF 0.50 0.50 0.50
Phi 59.86 59.62 59.94
f[Hz] 50.00
Error[%] 0.327
St.Dev[%] 0.003
LOADPOINT 2
U[V] 230.15 229.71 229.99
I[A] 40.03 40.08 40.03
P[W] 4651.26 4663.75 4625.82
Q[var] 7946.83 7940.81 7962.55
S[VA] 9206.78 9208.63 9209.09
PF 0.50 0.50 0.50
Phi 59.65 59.57 59.84
f[Hz] 50.00
Error[%] -0.158
St.Dev[%] 0.004