Working Standard
User’s Guide
WS-UGFW619-EN
www.appliedp.com
88
MEAS:CURR:AC:PHAS?
Request for phases between current and voltage in each phase.
MEAS:CURR:AC:SUM:SQ?
Request for sum of square currents.
MEAS:CURR:AC:DIST?
Request for the distortion of measured current.
MEAS:CURR:AC:TR:RAT
Transformer Test: Request for the measured transmission ratio of separate current channels (
r
Lx/L1
).
MEAS:CURR:AC:TR:ERR
Transformer Test: Request for the calculated transmission error of separate current channels (
er
Lx/L1
).
MEAS:CURR:AC:TR:PHAS
Transformer Test: Request for the measured phase error of separate current channels (
φ
Lx/L1
).
MEAS:CURR:AC:TR:RNOM
Transformer Test: Request for the nominal transmission current ratio.
MEAS:PHAS:PRES:RANG {<range>}
Setting the range: 0 = 0-360, 1 = +-180
MEAS:PHAS:PRES:RANG?
Request for phase range.
MEAS:PHAS:PRES:DIR {<range>}
Setting the range: 0 = Clockwise, 1 = Anticlockwise
MEAS:PHAS:PRES:DIR?
Request for phase direction
MEAS:PHAS:PRES:REL {<range>}
Setting the range: 0 = Relative, 1 = Absolute
MEAS:PHAS:PRES:REL?
Request for phase relation
MEAS:POW:AC[:ACT]?
Request for active power values.
MEAS:POW:AC[:ACT]:PRIM?
Request for primary active power values.
MEAS:POW:AC:REAC?
Request for reactive power values.
MEAS:POW:AC:REAC:PRIM?
Request for primary reactive power values.
MEAS:POW:AC:APP?
Request for apparent power values.