CSP-SNv Capacitor Charging Unit Operation
Manual
CSP-CSPSNV-8000/3
_______________________________________________________________________________________________
__________________________________________________________________________________________________________
Page
36
of
56
Local / Remote
Required:
CSP-CSPDV-7060 Remote Unit
CSP-CSPDV-4060 Remote Cable
The Local / Remote setting allows the use of a Remote Unit for operation from a
laboratory or instrument room.
The Remote Unit mirrors the CSP HV ON and HV OFF switch indicators, plus EOC, KEY and
general FAULT indicators, regardless of Local / Remote setting. HV ON and HV OFF
switches are recessed to prevent accidental operation.
The LOCAL / REMOTE setting determines operation of the Remote Unit when connected:
Local / Remote set to Remote
•
HV can only be enabled from the Remote Unit, NOT from the CSP.
Local / Remote set to Local
•
HV can only be enabled from the CSP, NOT from the Remote Unit.
HV can only be activated with the CSP in OPERATION MODE.
The HV OFF button is operational on the CSP AND the Remote Unit
regardless of Local / Remote setting.
With Local / Remote set to Remote, a fault mode will persist if the Remote
Unit is not detected, e.g. unplugging Remote Unit during operation.
Key In & Key Out is functionally identical on the CSP and the Remote Unit.
Note if using external Key In, connect only at CSP or Remote Unit, not both
simultaneously.