CSP-SNv Capacitor Charging Unit Operation
Manual
CSP-CSPSNV-8000/3
_______________________________________________________________________________________________
__________________________________________________________________________________________________________
Page
17
of
56
KEY Indicator
Synchronised to Key Out.
HV Indicator
Indicates HV is enabled on system
Local / Remote Indicator
LOCAL Operation indicated by LED ON.
REMOTE Operation indicated by LED OFF.
FAULT Indicator
It will illuminate during internal fault modes, see LCD display for fault identification.
You will need to press RESET once the fault condition has been cured.
INTERLOCK Indicator
Illuminated if an interlock condition is detected, please see main display for details of
interlock. You will need to press RESET once the interlock fault condition has been
cleared.
TIME OUT Indicator
For safety, the HV is switched OFF after a user selectable period without a key pulse
present. This also helps preserve the capacitors, as all pulse discharge capacitors are
not designed to sit powered up indefinitely.
HV FAULT Indicator
Indicates an HV Output over-current, open-circuit condition or HV charger module
fault.