CSP-SNv Capacitor Charging Unit Operation
Manual
CSP-CSPSNV-8000/3
_______________________________________________________________________________________________
__________________________________________________________________________________________________________
Page
18
of
56
Rear Panel Fuses
The rear panel contains 3 fuse holders, and these are the only things we recommend
the operator checks if he has not received factory training. The fuses are rated as
follows:-
F1
3A Antisurge 1¼"
(240V)
F2
250mA Antisurge 1¼"
(240V)
F3
3A Antisurge 1¼"
(240V)
F1 Protects 12v power supply
F2 Protects the interlock board
F3 Protects the line voltage powering the high voltage relays and the front panel fans.
Fan Filters
The rear panels of the units have 2 fan filters for the cooling air intake. These filters
should be periodically cleaned to ensure maximum air flow. This can be done with a
small stiff brush.
SAFETY NOTICE
All interlocks and safety features are doubled. For example, if the HV connector is
removed, the HV is shut OFF by the logic which controls the charger module AND the
high voltage is switched OFF by a separate relay. The interlock circuit operates from its
own isolated 12V supply and operates 2 relays, as well as the control electronics.
However, it cannot be stressed enough that the operator and all those who might come
into contact with this equipment treat it with extreme caution and should not take any
safety feature for granted.
Ensure the CSP-SNv is electrically isolated from the sound source prior to any
user adjustments to the cabling or source. Despite being heavily insulated, the
HV cable should not be touched or held when the unit is operating. The load
should always be in the water before operation.