19
Antti-Teollisuus Oy
Scale
01-2020
NOTE SUCCESSFUL DATA TRANSFER FROM THE METER DOES NOT REQUIRE
TYPING OF THE ENTIRE COMMAND STRING. IT MAY ALSO BE ACCOMPLISHED BY
USING THE PRINT-OUT REQUEST COMMAND UNDER THE PROGRAMMING MENU 2
PROGRAMMING MENU 8 - FORMATTING THE ANALOGUE OUTPUT
This menu is for converting the analogue signal from the meter suitable for the Logicontrol.
Using the analogue card enables transfer of weight data to the display of logicontrol.
Depress the PAR key on the scale once. While "PRO" is shown on the display, press the
F1^ button repeatedly until 8-Out is shown on the display. Move forward in the menu by
depressing the PAR key and depress the F1^ key to change the settings.
Setting the analogue card
Press
Param-
eter
Value
or sta-
tus
Description
Pro
No
Moving to programming mode (
PRO
) or
returning to start mode (
NO
)
8X
Pro
8-Ovt
Selecting settings for output card(
8-Ovt
)
type
4-20
Determine the output signal type from
the analogue card. (4-20mA, 0-10V,
0-20mA)
asIN
1NP
Determine the output signal value - se-
lect instantaneous value on display. (inp,
hi, lo, tot)
aN-LO
0
Scaling the lower limit for output signal.
AN-HI
80000
Scaling the upper limit for output signal.
Change the value to 80000 using the
buttons
and
.
vdt
2.0
Refreshing rate for output signal. Change
the value to 2.0s using the buttons
and
.
Pro
NO
Moving to programming mode (
PRO
) or
returning to start mode (
NO
)
EnD
Exiting the programming mode