NOTE: If the loss is not known, Option 21, Transmission Measurement, or the
Offset Calibration can be used to measure the loss of the system.
Displaying Power in dBm and Watts
Step 8.
Press the
Units
soft key to display power in Watts.
Displaying Relative Power
Step 9.
With the desired base power level input to the Cell Master, press the
REL
soft
key. The message area will show
Relative: On
and the power reading will indi-
cate 100%.
If measuring a 10 dBm signal with
Rel
selected and the power level drops to 7
dBm, the Relative setting will show -3 dB or 50%, depending on the units set-
ting selected.
Step 10.
Press the
Units
soft key to display power in dBm. Since Relative is ON, the
power reading will be in dBr, relative to the base power level.
7-2
Chapter 7 Power Monitor
CELL MASTER
SYNTHESIZER
RF DETECTOR
HOLD
RUN
START
CAL
AUTO
SCALE
SAVE
SETUP
RECALL
SETUP
LIMIT
MARKER
SAVE
DISPLAY
RECALL
DISPLAY
MODE
FREQ/DIST
AMPLITUDE
SWEEP
SYS
ENTER
CLEAR
ESCAPE
ON
OFF
/
1
2
4
5
6
7
8
9
0
3
+
-
.
Site Master S820C
M o d u l a t i o n
L i n e
O p e r a t e
S t a n d b y
L e v e l
F r e q u e n c y
S y s t e m
O u t p u t
O n O f f
R F O u t p u t
5 0
9
C l e a r
E n t r y
B a c k
S p a c e
7
8
9
4
5
6
1
2
3
0
.
+ / -
RF
DET
PORT
(Option 5)
Figure 7-1.
MT8212B Cell Master Power Monitor Setup
Содержание Cell Master MT8212B
Страница 3: ......
Страница 5: ......
Страница 222: ...12 4 Chapter 12 GPS Feature Figure 12 3 Saved GPS Location Information...
Страница 231: ...Chapter 13 CDMA Measurement 13 9 Figure 13 5 CDMA Display with Markers...
Страница 255: ...Chapter 18 EVDO Mode 18 5...