3-3
Before Use
3
3.1
Installation
3.1.3
Power Source Voltage and Frequency
Supply the MT9812B with power in the range between 85 and 132 Vac or between
170 and 250 Vac of voltage, and between 47.5 and 63 Hz of frequency.
Although switching the setting between 100 and 200 V systems is not required.
WARNING
To replace the fuse, first turn the power source OFF
and pull the power source cord out from the power re-
ceptacle, and then, replace the fuse. Attempting to re-
place the fuse without pulling the power source cord out
from the power receptacle may cause electric shock.
Moreover, check whether the fuse to replace is the same
as that prescribed in the manual, or use a fuse of the
same rating or the same properties of the fuse indicated
on the back of the MT9812B frame. Using a wrong fuse
may cause frequent fusing, burnout or fires.
Содержание MT9812B
Страница 11: ...For Safety xi Laser Radiation Markings D E C A B G F 1...
Страница 20: ...xx...
Страница 24: ...IV...
Страница 38: ......
Страница 52: ...Section 3 Before Use 3 14...
Страница 68: ...Section 4 Operation 4 16...
Страница 118: ...Section 5 Remote Control 5 50...
Страница 135: ...App i Appendix Appendix Appendix A Performance Test Result Recording List A 1 Appendix B Error Code B 1...
Страница 136: ...Appendix App ii...
Страница 142: ...Appendix A Performance Test Result Recording List A 6...
Страница 144: ...Appendix B Error Code B 2...