Calibration
Procedure
Perform the following steps:
Equipment Required:
Refer to Table 1-1 for further information about the following equip-
ment:
q
Anritsu ML24XXA with Version 2.02 or later, with Anritsu Power
Sensors MA2474A (use sensor SC6230 for 50 and 65 GHz mod-
els)
q
GPIB cable (Anritsu 2100-2, or equivalent)
NOTE
Allow the VNA and power meter to warm-up for at least 30
minutes prior to performing this calibration.
Procedure:
Step 1.
Press the Option Menu key (Enhancement key group) to
display the OPTIONS menu.
Step 2.
Select DIAGNOSTICS from menu, then select in
sequence: H/W CALIBRATIONS | SOURCE ALC CALI-
BRATION (next page).
ADJUSTMENTS
RF POWER/ALC CALIBRATION
37XXXD MM
6-11