Step 7.
Using the Marker Menu and Readout Marker key menus,
record the absolute value of markers 1 and 2 as follows:
q
Subtract one from the other.
q
Halve the difference
q
Add the resultant to the value of the marker at the low-
est peak (or the deepest trough).
This is the average value of the two peaks (or troughs).
Step 8.
Record the marker 3 value.
Step 9.
Find the absolute difference of the values recorded in step 7
and step 8. This is the Peak-to-Peak Ripple value. Use the
RF Measurement Chart (Table D-3) to find the correspond-
ing Return Loss value. This value is the Effective Source
Match. Insure Source Match measures at least 30 dB.
Step 10.
Change the S Parameter to S22. Move the outside wave-
guide section with the short attached to the Port 2 module.
Step 11.
Repeat Steps 4 through 9 to compute Port 2 Source Match.
Verify Port 2 Source Match is 30 dB minimum.
APPENDIX D
PERFORMANCE AND OPERATIONAL TESTS
37XXXD MM
D-17
Figure D-2.
Markers
Содержание 37 D Series
Страница 3: ......
Страница 4: ......
Страница 12: ......
Страница 14: ...1 2 37XXXD MM Figure 1 1 Series 37XXXD Vector Network Analyzer System...
Страница 26: ......
Страница 35: ...REPLACEABLE PARTS PARTS ORDERING INFORMATION 37XXXD MM 2 9 2 10 Figure 2 2 Signal Source Parts Location Diagram...
Страница 36: ......
Страница 38: ......
Страница 62: ......
Страница 80: ......
Страница 82: ......
Страница 90: ......
Страница 104: ......
Страница 122: ......
Страница 143: ...REMOVE AND REPLACE SIGNAL SOURCE MODULES 37XXXD MM 8 23 Figure 8 11 Signal Source Module Assemblies Removal Details...
Страница 148: ......
Страница 152: ......
Страница 154: ......
Страница 156: ......
Страница 158: ......
Страница 160: ...D 4 37XXXD MM ME7808 BLOCK DIAGRAM ME7808 BROADBAND SYSTEM MAINTENANCE Figure D 1 ME7808B C Block Diagram...
Страница 184: ......
Страница 190: ......