TROUBLESHOOTING
NON-RESPONSE OR OTHER UNUSUAL BEHAVIOR
37XXXD MM
7-13
Fails in
Forward only ?
Fails below 38
GHz?
Fails above 38
GHz?
Fails above 38
GHz?
Normal power at
Port 1 (S21)?
Normal power at
Port 1(S21)?
Normal power at
Port 1(S21)?
Sampler/
DownConversion
Module, A4 or A8
failure
A8, A4, or Sampler/
DownConversion
Module failure
transfer sw or A8
failure
not logical
Sampler/
DownConversion
Module, A4, or A8
failure
Ref A MUX coupler
failure
SPDT switch, Ref A
SQM, A8 or MUX
coupler failure
Yes
Yes
Yes
Yes
Yes
Yes
Yes
No
No
No
No
No
No
No
1
Go to
2
Figure 7-2.
Error Code DE Lock Failure Troubleshooting (50 and 65 GHz models only) (2 of 3)
Содержание 37 D Series
Страница 3: ......
Страница 4: ......
Страница 12: ......
Страница 14: ...1 2 37XXXD MM Figure 1 1 Series 37XXXD Vector Network Analyzer System...
Страница 26: ......
Страница 35: ...REPLACEABLE PARTS PARTS ORDERING INFORMATION 37XXXD MM 2 9 2 10 Figure 2 2 Signal Source Parts Location Diagram...
Страница 36: ......
Страница 38: ......
Страница 62: ......
Страница 80: ......
Страница 82: ......
Страница 90: ......
Страница 104: ......
Страница 122: ......
Страница 143: ...REMOVE AND REPLACE SIGNAL SOURCE MODULES 37XXXD MM 8 23 Figure 8 11 Signal Source Module Assemblies Removal Details...
Страница 148: ......
Страница 152: ......
Страница 154: ......
Страница 156: ......
Страница 158: ......
Страница 160: ...D 4 37XXXD MM ME7808 BLOCK DIAGRAM ME7808 BROADBAND SYSTEM MAINTENANCE Figure D 1 ME7808B C Block Diagram...
Страница 184: ......
Страница 190: ......