![Andor Technology New iStar Скачать руководство пользователя страница 115](http://html.mh-extra.com/html/andor-technology/new-istar/new-istar_user-manual_2942426115.webp)
Page 115
New iStar ICCD
6.2.2.1.2 - Integrate on Chip (IoC) enabled
6.2.2.1.2.1 - Default: one Pulse per trigger
Description
Several gate pulses can be generated within an exposure. A gate delay is applied that is the same per every trigger
within an exposure. Every external trigger within an exposure can generate only one gate pulse.
Setup
IOC enabled, output A/B/C disabled.
Waveforms
Operation
One Pulse per Trgger
External Trigger
Fire
gate Monitor
ARM