![Andor Technology New iStar Скачать руководство пользователя страница 101](http://html.mh-extra.com/html/andor-technology/new-istar/new-istar_user-manual_2942426101.webp)
Page 101
New iStar ICCD
Operation
6.1.3.2 - Number of accumulations > 1
A number of accumulated images will be acquired in the series.
6.1.3.2.1 - standard operation (IoC Disabled and Gate step Disabled)
Description
One gate pulse is generated within an exposure. A gate delay is applied that is the same for each exposure within an
accumulated image. The gate delay is also the same for every accumulated image that is returned.
Setup
IOC disabled, gate step disabled, Output A enabled.
Waveform
Fire
gate Monitor
Output A