![Andor Technology New iStar Скачать руководство пользователя страница 108](http://html.mh-extra.com/html/andor-technology/new-istar/new-istar_user-manual_2942426108.webp)
Page 108
New iStar ICCD
6.1.5.3 - Gate step enabled
Description
One gate pulse is generated within an exposure. For every successive exposure in the series, the gate delay applied is
incremented by the gate step value.
The fire pulse rate cannot exceed 15 kHz in this particular mode.
Setup
Gate step enabled, Output A enabled (trackstep disabled).
Waveforms
Operation
Fire
gate Monitor
Output A